Página 1 - Quick-Check

Quick-CheckTension MeterUser Instructions43079-0014 Issue AE April 2013

Página 2

10 Quick-Check User Instructions2.3 Important FeaturesQuick to use Attaches and removes from tensioned line in seconds. Quick-tensioning readout f

Página 3 - Table of Contents

Quick-Check User Instructions 11 3 OperationTypical operation of the Quick-Check is covered below, followed by explanations of the various display m

Página 4

12 Quick-Check User Instructions4. Release the lever arm and you are ready to perform another measurement.Figure 3.2 Quick-Check attached to cable3.

Página 5 - 1.1 About this manual

Quick-Check User Instructions 13 The Quick-Check has an internal temperature sensor inside the electronics cavity. Dramatic temperature changes (suc

Página 6 - 1.5 Sharp objects

14 Quick-Check User InstructionsYou can enter the temperature in one of two ways; let the Quick-Check determine the ambient temperature automatically

Página 7 - European Countries

Quick-Check User Instructions 15 Temp Press this softkey to choose the source of the temperature reading, the Quick-Check itself (Meter), outside in

Página 8 - 2 Introduction

16 Quick-Check User InstructionsThere is a keypad shortcut for increasing and decreasing contrast. While in normal display mode press the Arrow key a

Página 9 - 2.2 Front Panel and Keys

Quick-Check User Instructions 17 4 Configuration Mode4.1 Accessing the Configuration ModeYou need to access the Configuration mode to perform certa

Página 10 - 2.3 Important Features

18 Quick-Check User Instructions4. The unit is now in the Configuration mode. To see the rest of the softkeys available in this mode, press the Right

Página 11 - 3 Operation

Quick-Check User Instructions 19 Enable lbf - Enable or disable the pound-force unit of measure.Enable kgf - Enable or disable the kilogram-force un

Página 12 - 3.2 Measurement Practices

quickcheck_u_en_43079_0014.book© Avery Weigh-Tronix, LLC 2013. All rights reserved. No part of this publication may be reproduced, stored in an electr

Página 13 - 3.3 Softkey Functions

20 Quick-Check User Instructions5 Changing SheavesDo not use the Quick-Check with cable larger than indicated on the sheaves. Overload and damage to

Página 14

Quick-Check User Instructions 21 6 Achieving Best Accuracy6.1 AccuracyThe Quick-Check is an instrument designed to give accuracy that typically exc

Página 15

22 Quick-Check User Instructions6.5 Non-linearityMost three-point tension meters employ only linear characterization and have large errors at the mi

Página 16

Quick-Check User Instructions 23 7 TroubleshootingProblem Possible Cause SolutionPowers on momentarily and turns offLow battery Replace with high qu

Página 17 - 4 Configuration Mode

24 Quick-Check User Instructions8 SpecificationsPower2 AA, common alkaline batteries. Operational battery life over 40 hours at room temperature. Bat

Página 18

Quick-Check User Instructions 25 Available optionsVaried wire sizesOperating environmentSuitable for outdoor useDimensions10˝ x 23˝ x 3˝ (25 cm x 59

Página 19

26 Quick-Check User Instructions

Página 21 - 6 Achieving Best Accuracy

Overload Protection and Overhead Weighing EquipmentFairmont, Minnesota U.S.A.Toll-Free: (800) 368-2031Phone: (507) 238-4461Fax: (507) 238-8258www.dill

Página 22 - 6.5 Non-linearity

Quick-Check User Instructions 3 Table of ContentspageChapter 1 General information and warnings ...

Página 23 - 7 Troubleshooting

4 Quick-Check User Instructions

Página 24 - 8 Specifications

Quick-Check User Instructions 5 1 General information and warnings1.1 About this manualThis manual is divided into chapters by the chapter number a

Página 25 - Dimensions

6 Quick-Check User Instructions1.1.3 Safe handling of equipment with batteriesCAUTION: Danger of explosion if battery is incorrectly replaced. Repla

Página 26

Quick-Check User Instructions 7 1.6 FCC and EMC declarations of complianceUnited StatesThis equipment has been tested and found to comply with the

Página 27

8 Quick-Check User Instructions2 IntroductionThis manual covers the setup and operation of the Quick-Check Clamp Line Tensiometer from Dillon. The Qu

Página 28

Quick-Check User Instructions 9 2.2 Front Panel and KeysThe front panel of the Quick-Check is shown in Figure 2.2.Figure 2.2 Quick-Check front pan

Comentários a estes Manuais

Sem comentários