Dillon EDjunior (Red) Manual do Utilizador Página 20

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 19
18 EDjunior Dynamometer User Instructions
7 Troubleshooting
Problem Possible Cause Solution
EDjr powers on momentarily
and turns off
Low battery
Bad keypad
Replace with high quality alkaline batteries
Have EDjr serviced
EDjr does not power on Low battery
Bad keypad
Replace with high quality alkaline batteries
Have EDjr serviced
Batteries installed backwards
or no spring contact
Insure that positive terminals of both batteries (nub) face inward –
towards the black cap. Check that spring is attached to the battery
cap.
Software reset Remove battery cap & reinstall after one minute. Attempt to turn
power on again.
Display contrast too light Hold the Arrow key down while pressing the F2 key several times to
increase the display contrast. If nothing occurs, release both keys.
Press the power button and try again.
Display is completely dark Display contrast too dark Hold the Arrow key down while pressing the F1 key several times to
decrease the display contrast.
Display flashes Low battery Replace with high quality alkaline batteries